A system has been developed for measuring non-shunt cardiac output by the throughflow technique, using nitrous oxide in patients undergoing general anaesthesia. The throughflow measurement technique is a non-invasive method based on inert gas throughflow theory. In vitro validation of the measurement system was performed using a lung gas exchange simulator. The accuracy and precision of the throughflow measurement system was assessed by comparing measured and target values for five simulated values of non-shunt cardiac output, from 2.88 to 9.86 l min(-1). This showed an overall mean bias of -0.031 min(-1) (range -0.00 to -0.101 min(-1)), with a mean coefficient of variation of the difference of 1.39% (1.20-1.93%). These results indicate that the measurement system is suitable for monitoring the non-shunt cardiac output in patients undergoing general anaesthesia using nitrous oxide throughflow.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.