Background-The majority of prosthetic heart valves currently implanted are tissue valves that can be expected to degenerate with time and eventually fail. Repeat cardiac surgery to replace these valves is associated with significant morbidity and mortality. Transcatheter heart valve implantation within a failed bioprosthesis, a "valve-in-valve" procedure, may offer a less invasive alternative. Methods and Results-Valve-in-valve implantations were performed in 24 high-risk patients. Failed valves were aortic (nϭ10), mitral (nϭ7), pulmonary (nϭ6), or tricuspid (nϭ1) bioprostheses. Implantation was successful with immediate restoration of satisfactory valve function in all but 1 patient. No patient had more than mild regurgitation after implantation. No patients died during the procedure. Thirty-day mortality was 4.2%. Mortality was related primarily to learning-curve issues early in this high-risk experience. At baseline, 88% of patients were in New York Heart Association functional class III or IV; at the last follow-up, 88% of patients were in class I or II. At a median follow-up of 135 days (interquartile range, 46 to 254 days) and a maximum follow-up of 1045 days, 91.7% of patients remained alive with satisfactory valve function. Conclusions-Transcatheter valve-in-valve implantation is a reproducible option for the management of bioprosthetic valve failure. Aortic, pulmonary, mitral, and tricuspid tissue valves were amenable to this approach. This finding may have important implications with regard to valve replacement in high-risk patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.