A comparison of electric field measurements was carried out between January 2010 and July 2012 amongst fourteen national metrology laboratories at frequencies 1 GHz, 2.45 GHz, 10 GHz and 18 GHz at indicated field levels of 10 V/m, 30 V/m and 100 V/m. The identification of this intercomparison is CCEM.RF-K24.F. Two electric field probes provided by Amplifier Research have been measured as the travelling standards. The National Physical Laboratory (United Kingdom) acted as the pilot laboratory for the comparison. This report contains the results of the measurements.Main text.
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
We consider the results of key comparisons of standards of electric field intensity in the frequency range 1-18 GHz. As a result of these comparisons, the data presented in the BIPM database confirmed possibilities for calibrations and measurements of electric field intensity at the indicated frequencies at VNIIFTRI.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.