A simple, sensitive and specific method for the quantification of N-acetyl-S-(N-methylcarbamoyl)cysteine (AMCC) in urines of the general population and of exposed workers has been developed. Samples were first diluted with phosphate buffer followed by purification by solid-phase extraction (SPE) with SAX columns prior to analysis by liquid chromatography/mass spectrometry (LC/MS). An external standard was used for the quantification together with selected ion monitoring of m/z 219, [M-H]-. The linear calibration curve showed a good correlation (r2 = 0.98, p < 0.001) between 0.1-110 mg/L of AMCC, and the detection limit was calculated to be 2 microg/L. The within-day and between-day precision, calculated for exposed workers and general population AMCC levels in urine samples, were determined as 1.2 and 3.6%, respectively (as relative standard deviation (RSD)).
Due to the good agreement between the two groups of data, the TK200 samplers can be considered as a simpler approach than the pump for screening worker exposures to DMF.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.