There is a growing interest in Gafchromic films for patient dosimetry in radiotherapy and in radiology. A new model (XR-QA) with high sensitivity to low dose was tested in this study. The response of the film to different x-ray beam energies (range 28-145 kVp with various filtrations, dose range 0-100 mGy) and to visible light was investigated, together with the after exposure darkening properties. Exposed films were digitized with a commercially available, optical flatbed scanner. A single functional form for dose versus net pixel value variation has been determined for all the obtained calibration curves, with a unique fit parameter different for each of the used x-ray beams. The film response was dependent on beam energy, with higher colour variations for the beams in the range 80-140 kVp. Different sources of uncertainties in dose measurements, governed by the digitalization process, the film response uniformity and the calibration curve fit procedure, have been considered. The overall one-sigma dose measurement uncertainty depended on the beam energy and decreased with increasing absorbed dose. For doses above 10 mGy and beam energies in the range 80-140 kVp the total uncertainty was less than 5%, whereas for the 28 kVp beam the total uncertainty at 10 mGy was about 10%. The post-exposure colour variation was not negligible in the first 24 h after the exposure, with a consequent increase in the calculated dose of about 10%. Results of the analysis of the sensitivity to visible light indicated that a short exposure of this film to ambient and scanner light during the measurements will not have a significant impact on the radiation dosimetry.
Gafchromic XR-R films are a useful tool to evaluate entrance skin dose in interventional radiology. Another dosimetric quantity of interest in diagnostic and interventional radiology is the dose area product (DAP). In this study, a method to evaluate DAP using Gafchromic XR-R films and a flat-bed scanner was developed and tested. Film samples were exposed to an x-ray beam of 80 kVp over a dose range of 0-10 Gy. DAP measurements with films were obtained from the digitalization of a film sample positioned over the x-ray beam window during the exposure. DAP values obtained with this method were compared for 23 cardiological interventional procedures with DAP values displayed by the equipment. The overall one-sigma dose measurement uncertainty depended on the absorbed dose, with values below 6% for doses above 1 Gy. A maximum discrepancy of 16% was found, which is of the order of the differences in the DAP measurements that may occur with different calibration procedures. Based on the results presented, after an accurate calibration procedure and a thorough inspection of the relationship between the actual dose and the direct measured quantity (net optical density or net pixel value variation), Gafchromic XR-R films can be used to assess the DAP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.