A 24 mm COMS‐like eye plaque was developed to meet the treatment needs of our eye plaque brachytherapy practice. As part of commissioning, it was necessary to determine the new plaque's seed coordinates. The FARO Edge, a commercially available measurement arm, was chosen for this purpose. In order to validate the FARO Edge method, it was first used to measure the seed marker coordinates in the silastic molds for the standard 10, 18, and 20 mm COMS plaques, and the results were compared with the standard published Task Group 129 coordinates by a nonlinear least squares match in MATLAB version R2013a. All measured coordinates were within 0.60 mm, and root mean square deviation was 0.12, 0.23, and 0.35 mm for the 10, 18, and 20 mm molds, respectively. The FARO Edge was then used to measure the seed marker locations in the new 24 mm silastic mold. Those values were compared to the manufacturing specification coordinates and were found to demonstrate good agreement, with a maximum deviation of 0.56 mm and a root mean square deviation of 0.37 mm. The FARO Edge is deemed to be a reliable method for determining seed coordinates for COMS silastics, and the seed coordinates for the new 24 mm plaque are presented.PACS number: 87.53.Jw
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.