This paper describes the establishment by the Institute of Nuclear Energy Research (INER, Taiwan) of the reference air kerma rate (RAKR) calibration standard for measurement of high-dose rate (HDR) (192)Ir brachytherapy source strength. A bilateral comparison has been made in the RAKR standards for HDR (192)Ir brachytherapy sources at the INER and Physikalisch-Technische Bundesanstalt (PTB, Germany) and the measurement difference was within the overall standard uncertainty and showed good agreement between the two calibration standard systems established at the INER and the PTB. Besides, INER also worked with 20 domestic hospitals to organise an on-site measurement comparison programme to explore the status of HDR (192)Ir brachytherapy source strength determination in Taiwan. The comparison results presented the ratios of RAKR with vendor values, as determined by INER and hospitals from the programme. The ratios fall in all cases within the ± 3 % guaranteed by the vendors for a coverage factor of k = 2 or at 95 % confidence level.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.