Trace elements in textile fibers were analyzed by the in-air particle induced X-ray emission (PIXE) method. Individual fibers were used as a target, representing minute material available in forensic or archeological applications. For quantification of the concentrations, the X-ray production was assumed to occur in a thin cylindrical target that intersects the central part of the beam. The irradiated volume and the distribution of escape lengths were calculated in the model. The impact beam intensity was axially symmetric Gaussian, as shown by calculation and cross-scan measurements. The normalization procedure relied on the argon signal from the air, which was checked by the chopper method. For a set of fibers, elements characteristic of the fibers and their dressings were determined.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.