We assessed the usefulness of PET/CT images to determine the target volume in radiotherapy planning by evaluating the standardized uptake value (SUV). We evaluated the imaging conditions and image-reconstruction conditions of PET/CT useful for treatment planning by collecting (18)F-FDG images of acrylic spheres (10-48 mm in diameter) in a phantom. The (18)F-FDG concentration in the spheres was 10-fold higher than that of the phantom. The contours were delineated according to the SUV by the threshold and regions of interest (ROI) methods. Comparisons of two- and three-dimensional (2D and 3D) acquisition images indicated that the sharpness and quantitative qualities of the sphere boundaries were better in the former than in the latter. In the extraction of outlines using the SUV, outlines obtained at an SUV of 40-50% of the maximum agreed well with the actual acrylic sphere size. 2D acquisition images are more suitable for delineating target volume contours, although 3D acquisition images are more popular in diagnostic imaging. An SUV of 40-50% of the maximum is suggested to be appropriate for GTV contouring of sphere tumors with homogenously distributed (18)F-FDG.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.