µDose is a novel compact analytical instrument for assessing low level 238 U, 235 U, 232 Th decay chains and 40 K radioactivity. The system is equipped with a dual α/β scintillator allowing discrimination between α and β particles. The unique built-in pulse analyzer measures the amplitude of each individual pulse, its shape and the time interval between subsequent pulses. This allows the detection of pulse pairs arising from subsequent decays of 214 Bi/ 214 Po, 220 Rn/ 216 Po, 212 Bi/ 212 Po and 219 Rn/ 215 Po. The obtained α and β counts and four separate decay pair counts are used to calculate 238 U, 235 U, 232 Th and 40 K specific activities in measured samples through the use of radioactivity standards. The µDose system may be equipped with various photomultipliers and counting containers to assess radionuclide concentrations of samples of masses ranging between 0.4 g and 4 g. As a result, the user can customize the system to their needs and maximize the instrument's performance. The system is controlled by dedicated software with a graphical user interface and modules for system calibration, data visualization, specific radioactivity calculations and dose rate determination using the infinite matrix assumption.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.