Intracoronary OCT appears to be feasible and safe. Optical coherence tomography identified most architectural features detected by IVUS and may provide additional detailed structural information.
Background-We report findings from optical coherence tomography (OCT) of in-stent neoatherosclerosis as a cause of drug-eluting stent (DES) failure.
Methods and Results-Optical
024).Compared with DES Ͻ20 months after implantation (the best cut-off to predict TCFA-containing neointima), DES Ն20 months after implantation had a higher incidence of TCFA-containing neointima (69% versus 33%, Pϭ0.012) and red thrombi (27% versus 0%, Pϭ0.007). Patients with unstable (versus stable) angina had an increasing number of unstable OCT findings including TCFA-containing neointima, neointima rupture, and thrombus (Pϭ0.027). The rate of agreement between grayscale intravascular ultrasound and OCT for detecting intimal rupture was 50% and for detecting thrombus was 44%. The agreement between virtual histology intravascular ultrasound and OCT for identifying TCFA-containing neointima was 78%. Conclusions-In-stent neoatherosclerosis may be an important mechanism of DES failure, especially late after implantation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.