Background
Continuous curvilinear capsulorhexis (CCC) is a delicate ophthalmic procedure which may benefit from robot technology. Measuring the behaviours (physiological tremor, operation force) of surgeons provides baseline data to develop assistive CCC robot.
Methods
A forceps with fibre bragg grating and inertial sensors is used to measure the surgeons' behaviours while experts/novices perform CCC on ex‐vivo pig eyes, in‐vivo rabbit eyes and ex‐vivo human lens.
Results
In pig/rabbit tests, the root‐mean‐square (RMS) tremor amplitude is 35.26/59.04 μm (expert/novice, transverse), 13.3/20.55 μm (axial). The RMS voluntary force (VF) and involuntary force (IF) are 8.97/17.16 mN, and 0.66/1.90 mN, respectively. In human lens test, the RMS tremor amplitude is 24.0 μm (transverse, expert only), 9.88 μm (axial). The RMS VF and RMS IF are 9.04 mN (expert only) and 0.17 mN, respectively.
Conclusions
The expert surgeons have better precision and less operation force.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.