Porcine C-peptide concentration can be reliably measured in porcine serum samples with either assay (ELISA or RIA). However, the C-peptide results generated by these two assays are not equivalent. Therefore, assay bias must be considered before directly comparing pre-clinical studies that used either of these assays. We determined that harmonization between the assays is appropriate in a specific concentration range. Outside of that range, we do not know whether a linear correction function can be more broadly applied. The variation between the two assays may be related to calibration or reagent factors. To determine which assay is truly more accurate and to effectively compare interlaboratory results, we will need a traceable reference standard.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.