Intraoperative fractures are a rare complication in total knee arthroplasty. Limited literature exists in regard to the incidence, mechanism of injury and management of intraoperative fractures. The authors report a unique case of an 80-year-old man who sustained a medial tibial plateau fracture that occurred intraoperatively during final tibia bone preparation with the use of the Woolley Tibia Punch (Innomed, Savannah, Georgia, USA). The fracture was managed with the addition of 4.5 mm cortical lag screws and the addition of a stemmed tibial implant to bypass the fracture. This is the first reported case in literature that describes an intraoperative medial tibial plateau that occurred through the use of a Woolley Tibia Punch. The authors recommend the consideration of drilling to prepare sclerotic bone for cement penetration rather than a punch in order to minimise the potential for intraoperative fractures that may occur with the use of a punch.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.