The transferability of the candidate Reference Method for total serum protein was tested in eight laboratories in the United States and Europe. National Bureau of Standards SRM 927 (bovine serum albumin) was used in each analytical run as the calibration standard. The mean absorptivity value obtained for this material was 0.2983 L g-1 cm-1. Four serum pools prepared at the Centers for Disease Control were analyzed on each of 15 days. Within-run variation of the protein values (expressed as CV) in the eight laboratories ranged from 0.1 to 2.5% and day-to-day (total) variation in six of the laboratories ranged from 0.4 to 1%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.