Background and purpose After total hip arthroplasty, dislocations are a frequent complication and are difficult to treat in some patients. A great variety of implants and antiluxation mechanisms are used in surgical therapy.Methods 8 patients had 9 Duraloc Constrained Inlays implanted at our clinic between October 2003 and November 2006, for recurrent dislocations. A retrospective follow-up study was carried out.Results All patients suffered a failure of the expanding ring, the metal ring being squeezed out of the polyethylene notch. The mechanism of failure can be explained by impingement due to the implant design. At the time of writing, 3 patients have had to undergo revision surgery.Interpretation The Duraloc Constrained Inlay has shown unacceptably high failure rates.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.