The INTRABEAM 50 kV x-ray device can be used for intra-operative partial breast irradiation. Spherical applicators are added to the source probe to deliver a radially symmetric radiation dose. Dosimetric data for calculation of absorbed dose were measured for this unit and a superficial unit with a similar beam quality, as defined by half value layer (HVL). Chamber calibration factors, N(K), and chamber correction factors, k(ch), were determined based on HVL, according to the IPEMB code of practice and addendum. Depth doses were also measured using an ionization chamber and GafChromic EBT film. HVL was measured as 0.85-1.30 mm Al across the range of applicator sizes. Values for N(K) and k(ch) were found to be similar for the two units and all INTRABEAM applicator sizes. Therefore, calibration of ionization chambers, radiochromic film and other relative dosimeters could be performed on the superficial unit. This has the advantage of higher dose rates and lower dependence on small variations in detector positioning. Depth dose measurements performed using film also agreed with chamber values, published and manufacturer data, giving a simple and robust method for commissioning and regular quality assurance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.