The STORZ TRUE VISTA bifocal intraocular lens was evaluated in a laboratory setting and during clinical trials. The Modulation Transfer Function and Through Focus Response were measured in a simulated eye chamber for selection of optimal add zone size, location, power and pupil dependency. Results indicate the optimaldesign to be an annular add power of+4D having a 1 .5 millimeterinner diameter and a 2.6 millimeter outer diameter. Optical performance was predicted using computer ray trace and Fourier analysis. Laboratory testing confirmed design performance achieved through the manufacturing process. Clinical results correlate well with Through Focus Response but less well with Modulation Transfer Function. 0-8194-11
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.