Recently, along with technology development of endoscopic equipment, the stent technology has been developed for the convenience of operation, shortening of recovery times, and reduction of patient's pain. In this study, paclitaxel-eluting metal stents for treatment of biliary benign stenosis were developed through an electrospray-coating method. Polyether-based polyurethane (PELLETHANE 2363-80AE ® ) and paclitaxel were coated onto the surface of a metallic stent and Pluronic F127 was used as an additive. As a result, physicochemical characterization of paclitaxel via SEM, FTIR, contact angle and XRD techniques revealed the information of solid state of paclitaxel-loaded PU film. The in vitro release profile showed a slower release rate with a higher content of paclitaxel.Keywords: stent, electrospray, pellethane, paclitaxel, drug release.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.