a b s t r a c tThe mechanical behavior of a variety of cable architectures comprised of silver-cored wires was evaluated in uniaxial tension, and in cyclic strain-controlled fatigue with the use of a flex tester operated to provide fully reversed bending fatigue. The magnitude of cyclic strains imparted to each cable tested was controlled via the use of different diameter mandrels. Smaller diameter mandrels produced higher values of cyclic strain and lower fatigue life. Multiple samples were tested and analyzed via scanning electron microscopy. The fatigue results were analyzed via a Coffin-Manson-Basquin approach and compared to fatigue data obtained on 316LVM cables where testing was conducted in an identical manner. The effects of changes in wire diameter, cable architecture, and wire composition (i.e. silver-cored vs. 316LVM) are discussed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.