An immunoassay performed on a portable microfluidic device was evaluated for the determination of urinary albumin. An increase in absorbance at 500 nm resulting from immunoagglutination was monitored directly on the poly(dimethylsiloxane) (PDMS) microchip using a portable miniature fibre-optic spectrometer. A calibration curve was linear up to 10 mg L–1 (r2 = 0.993), with a detection limit of 0.81 mg L–1 (S/N = 3). The proposed system showed good precision, with relative standard deviations (RSDs) of 5.1%, when evaluated with 10 mg L–1 albumin (n = 10). Determination of urinary albumin with the proposed system gave results highly similar to those determined by the conventional spectrophotometric method using immunoturbidimetric detection (r2 = 0.995; n = 15).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.