We have developed models describing the sensitivity and sampling volume of various remote fiber-optic Raman probes—single-fiber, lensed, dual-fiber beveled-tip, dual-fiber flat-tipped, and multi-fiber flat-tipped. The models assume clear samples and incorporate radii, separation, bevel angle, and numerical aperture of the fibers; overlap geometry of illumination and excitation light cones; and refractive index of immersion medium. For the Raman spectra of solid samples in air, single-fiber and lensed probes are predicted to yield the highest Raman signal. Beveled probes should provide greater Raman signal strength than do flat-tipped probes because beveled probes can collect light from a restricted volume closer to the probe end. Although multiple collection fibers improve Raman signal strength, progressively distant concentric fiber rings contribute less and sample material further from the probe.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.