We investigated a prototype virtual-pinhole positron emission tomography (PET) system for small-animal imaging applications. The PET detector modules were made up of 1.3 mm lutetium-yttrium oxyorthosilicate (LYSO) arrays, and the insert detectors consisted of 0.6 mm pixelated cadmium zinc telluride (CdZnTe). To validate the imaging experiment, we did a Monte Carlo simulation for the virtual-pinhole PET (VP-PET) system in the Geant4 Application for Emission Tomography (GATE). For a point source of 22Na with a 0.5 mm diameter, the filtered back-projection algorithm-reconstructed PET image showed a resolution of 0.7 mm full-width-at-half-maximum. The system sensitivity was 0.46 cps/kBq at the center of the field view of the PET system with a source activity of 0.925 MBq and an energy window of 350 to 650 keV. A rod source phantom and a Derenzo phantom with 18F were also simulated to investigate the PET imaging ability. GATE simulation indicated that sources with 0.5 mm diameter could be clearly detected using 0.6 mm pixelated CdZnTe detectors as insert devices in a VP-PET system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.