Abstract:We describe a 'wet mirror' apparatus for cw cavity-enhanced absorption measurements with Bacteriochlorophyll a (BChla) in solution and show that it achieves the full sensitivity gain (≈ 2.3 × 10 4 ) afforded by the finesse (3.4 × 10 4 ) and loss distribution of our optical resonator. This result provides an important proof-of-principle demonstration for solution-phase cavity-enhanced spectroscopy; straightforward extrapolation to a system with state-of-the-art low-loss mirrors and shot-noise-limited performance indicates that single molecule sensitivity in liquids is within reach of current technology. With the probe laser locked to the cavity resonance, our instrument achieves a sensitivity ≈ 3.4 × 10 −8 / √ Hz (for a sample of length 1.75 mm) with 100 kHz bandwidth and can reliably detect sub-nM concentrations of BChla with 1 ms integration time.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.