. AbstructThe differential diagnostic capacity of a multi-directional sensing catheter with intraluminal objects was evaluated by fluorescence analysis following administration of a photosensitizer. The measurement catheter is 1.7 mm in diameter and is constructed with cross-sectional arrangement of 6 sensor bundle units and a centered guide wire lumen. Fluorescence was analyzed by polychrometer and amplified CCD camera with an image intensifier. The excitation light was 405nm wavelength obtained from an Xe-Hg lamp using a band pass filter. In vivo and in vitro fluorescence analysis of normal and atherosclerotic rabbit aorta was performed following intravenous administration of 5 mg/kg Hematoporphyrin Derivatives (HpD) 24 hours before. The diagnostic algorithm was based on the spectrum profile, a peak at 405nm and Plaque Index (PI=[F1630-(O.76F1600-i-O.24F1725)] /F1600) was identified for the fluorescence spectrum detected by each one of sensing units. Blood and fresh thrombus showed a flat profile without a 405nm peak (sens.=83.3%, spec.=100%). The normal arterial wall was PI O.15 with a 405nm peak. Intima of less than 200p.m was O.15
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.