Grazing incidence X-ray fluorescence analysis with a collodion film sample holder was performed by using a portable total reflection X-ray fluorescence spectrometer. The spectral background was low when using this sample holder, and a Cr detection limit obtained with this sample holder was comparable to that obtained with a diamond-like carbon coated quartz glass sample holder which was usually used as sample holder for the portable spectrometer. A river water sample was also measured with the collodion film sample holder, and 10.0 μg/L of Cr and 5.1 μg/L of Mn were detected. Collodion film sample holder can be used as a disposable sample holder for trace elemental analysis using the portable spectrometer.
Using a portable total-reflection X-ray fluorescence (TXRF) spectrometer with a collodion film sample holder, a spectrum of an analyte containing 50 ng of aluminum was measured. The Al Kα line (1.49 keV) that partially overlaps with the Si Kα line (1.74 keV) from a quartz glass substrate usually used as a sample holder for TXRF analysis, was clearly detected when using the collodion film sample holder. To investigate the quantitative performance of the portable spectrometer with a collodion film sample holder, the concentrations of Cr, Mn, and Fe in a certified reference material of river water (JSAC 0302-3b), whose certified values are 10.0, 5.1, and 59.6 μg/L, respectively, were determined by the internalstandard method. We showed that approximate concentrations of these elements were determined.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.