We present the methodology used to fabricate an X-ray reflection grating and describe a technique for grating replication. Further, we present the experimental procedure and results of a study to measure the diffraction efficiency of a replicated X-ray reflection grating in an extreme off-plane geometry. The blazed grating demonstrates a total diffraction efficiency of ∼60% from 0.34 to 1.2 keV at a grazing angle of ∼1.°5, with single-order efficiency ranging from ∼35% to 65% for energies within the blaze envelope. The diffraction efficiency of the grating measured relative to the reflectivity of the metal coating averages ∼90% above 0.34 keV. Data collected as a function of beam position on the grating indicate a relative variation in total efficiency of <1% rms across the grating surface.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.