A new ICD detects AT/AF accurately and continuously. Therapy may be programmed for long-duration AT/AF, with a low risk of underdetection. Discrimination of AT from AF permits successful pacing therapy for a significant fraction of AT.
The wavelet morphology algorithm has high sensitivity for VT detection. Inappropriate detections of SVT as VT may be reduced by optimal programming. Downloadable software permits evaluation of new algorithms in implanted ICDs.
In the canine model, measurement of chronic intrathoracic impedance with an implantable system effectively revealed changes in thoracic congestion due to heart failure reflected by LVEDP. These data suggest that implantable device-based impedance measurement merits further investigation as a tool to monitor the fluid status of heart failure patients.
Leakage current causes cardiovascular collapse at levels below the VF threshold. Stimulation by silent AC that is neither felt nor visible on the ECG presents as hypotensive VT. In patients with intracardiac electrodes, leakage current less than or equal to the present standard of 50 microA may cause VT or VF. The safety standard for leakage current lasting >/=5 seconds should be =20 microA. This standard should be based on the continuous-capture threshold.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.