The purpose of this study was to compare stiffness and resistance to fracture of four brands of nickel titanium K files. Instruments of sizes 15 to 40 were tested according to ANSI/ADA Specification No. 28. Resistance to fracture was determined by twisting and measuring the maximum torque and angular deflection at failure. Stiffness was determined by measuring the moment required to bend the instrument 45 degrees. The permanent deformation angle remaining between the tip and the flutes of the instruments after bending ceased was also recorded. Nickel titanium K files satisfied and far exceeded specification standards for stiffness. They also satisfied and exceeded the standards for angular deflection at failure. They met or exceeded the maximum torque at failure standards in all sizes except for the size 40 of the Maillefer Niti, and the size 30 of the MacSpadden Niti. Nickel titanium K files presented a null permanent deformation angle. Clinical studies are required to evaluate the influence of low bending moment on other properties such as breakage and canal transportation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.