This study aimed to compare the difference in mechanical behavior between two types of female-part clips of the Stress-Free Implant Bar (SFI-Bar) system during simulation of insertion-removal cycles. Materials and Methods: A total of 10 samples simulating SFI-Bar-attachment-retained implant overdentures were fabricated and randomly divided into two groups (n = 5). One group used E-clips (Elitor alloy) as the female part of the SFI-Bar, and the other used T-clips (all titanium grade IV with red nylon inserts). A total of 14,000 insertionremoval cycles were carried out on each sample. Retentive forces from each cycle were recorded for analysis. Results: Significant differences were found between the two groups (P < .05). Conclusion: The retentive force of E-clips increased as the number of dislodging cycles increased, suggesting that some adjustment may be needed to lower this part's retentive force. T-clips with changeable nylon inserts were deformed after about 4,200 insertion-removal cycles, which interfered with insertion. This indicated that T-clips may need replacement after 2 to 3 years of clinical use.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.