A miniature fiber Doppler imaging catheter for integrated functional and structural optical coherence tomography (OCT) imaging has been developed. The Doppler catheter can be use to map blood flow profile within a vessel as well as image vessel wall structures. The prototype Doppler catheter was demonstrated in measuring the intraluminal velocity profile in a vessel phantom (conduit). A simple mathematical model can be used to estimate the flow profile outside of normal OCT beam penetration. By extending the spatial range of the flow measurements to approximately two times the normal OCT image penetration depth, the total flow rate can then be calculated from the estimated velocity profiles. The measured total flow rate in the vessel phantom obtained from the Doppler imaging catheter correlates well with the calibrated flow values. The Doppler OCT catheter's ability to simultaneously obtain both structural and functional information makes it a potentially powerful device of cardiovascular imaging. 10 270 Proc. SPIE Vol. 4619 113 Downloaded From: http://proceedings.spiedigitallibrary.org/ on 06/16/2016 Terms of Use: http://spiedigitallibrary.org/ss/TermsOfUse.aspx
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.