Although the values of different parameters may remain within permissible limits during the operation of a thermoluminescent dosemeter (TLD) reader, certain effects can become apparent only when a long-term followup of these parameters is performed. In order to ensure an accurate and reliable operation of a TLD reader, the system characteristics must be monitored continuously. Long-term statistical checks of key system parameters may give a broader insight into the operational characteristics of the TLD reader and may help for proper maintenance of the system. The photomultipliers noise, the internal reference light source stability and the A to D reference voltage were found to be critical parameters, which have a major influence on the accuracy and stability of the system. A followup of these parameters for a period of about 10 y is presented, and some problems are seen to be reflected in the distributions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.