We evaluate the combined usage of two systems, the Integral Quality Monitor (IQM) transmission detector and SoftDiso software, for in vivo dose monitoring by simultaneous detection of delivery and patient setup errors in whole breast irradiation. An Alderson RANDO phantom was adapted with silicon breast prostheses to mimic the female anatomy. Plans with simulated delivery errors were created from a reference left breast plan, and patient setup errors were simulated by moving the phantom. Deviations from reference values recorded by both monitoring systems were measured for all plans and phantom positions. A 2D global gamma analysis was performed in SoftDiso for all phantom displacements. Both IQM signals and SoftDiso R-values are sensitive to small MU variations. However, only IQM is sensitive to jaw position variations. Conversely, IQM is unable to detect patient positioning errors, and the R-value has good sensitivity to phantom displacements. A gamma comparison analysis allows one to determine alert thresholds to detect phantom shifts or relatively large rotations. The combined use of the IQM and SoftDiso allows for fast identification of both delivery and setup errors and substantially reduces the impact of error identification and correction on the treatment workflow.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.