Human immunodeficiency virus type-1 (HIV-1) viral load is useful for monitoring disease progression in HIV-infected individuals. We generated RNA standards of HIV-1 and internal control (IC) by in vitro transcription and evaluated its performance in a quantitative reverse transcription polymerase chain reaction (qRT-PCR) assay. HIV-1 and IC standards were obtained at high RNA concentrations, without DNA contamination. When these transcripts were included as standards in a qRT-PCR assay, it was obtained a good accuracy (±0.5 log10 unit of the expected results) in the quantification of the HIV-1 RNA international standard and controls. The lower limit detection achieved using these standards was 511.0 IU/mL. A high correlation (r = 0.925) was obtained between the in-house qRT-PCR assay and the NucliSens easyQ HIV-1 test (bioMerieux) for HIV-1 RNA quantitation with clinical samples (N = 14). HIV-1 and IC RNA transcripts, generated in this study, proved to be useful as standards in an in-house qRT-PCR assay for determination of HIV-1 viral load.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.