In the report we consider the experimental set-up for measuring the main characteristics of X-ray capillary products in the X-ray energy band ranging from 4 keV up to 60 keV. These characteristic are: 1) the capillary/lens transmission coefficient; 2) X-ray focal spot sizes of lenses and conical capillaries; 3) focal distances of lenses; 4) the transmission coefficient and the scatter reduction factor (for anti scatter grids). The arrangement is designed to allow measuring spectral dependence of the above characteristics. Example of the measurement of the X-ray poly-capillary transmission coefficient is given. The influence of mechanical and electronic stability on the measured values is discussed. Examples of the designed and manufactured gimbal mounts, translation stages, X-ray collimators and X-ray detectors that were produced by IRO in collaboration with 'Kurchatov Institute' are presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.