any of the instruments used in this study. Flattest and steepest keratometric data and also IOP showed highly significant correlation with the ten specific CorVis ST parameters.
Conclusions:The novel technique measuring ocular biomechanical properties showed high repeatability for only IOP and pachymetric data determined in our study. Single measurements are not reliable in case of the ten device-specific parameters. The equipment allows for conducting clinical examinations, and screenings of surgeries altering ocular biomechanical properties with some form of averaging of multiple measurements.
Purpose: Our aim was to assess the corneal power, axis, and age dependence of the anterior and posterior corneal surfaces with Scheimpflug imaging.Methods: Patients older than 10 years without a history of ocular surgery, corneal diseases, contact lens wearing, or severe dry eye were enrolled. Data regarding the anterior and posterior cornea were analyzed by means of Scheimpflug imaging (Pentacam HR).Results: The median age was 46.8 years (range: 10.0-90.3), and involved 827 eyes of 827 patients. Posterior corneal astigmatism was a median of 20.30 diopters (D) and exceeded 0.50 D in 12.56%. The ratio of the anterior-to-posterior corneal radius was a median of 1.217 (range: 1.021-1.402). The correlation between this ratio and age was significant (r = 20.219; P , 0.001). The correlation between anterior and posterior corneal astigmatism and age was significant. At the anterior surface, the ratio of with-the-rule astigmatism was 80.0% for the age range 10 to 20, decreasing to 36.0% for those aged over 81 years, whereas the ratio of against-the-rule astigmatism increased from 7.1% to 44.0%. At the posterior surface, the ratio of with-the-rule astigmatism was 94.3% for the age range 10 to 20 years, decreasing to 84.0% for those aged over 81 years, whereas the ratio of against-the-rule astigmatism increased from 1.4% to 8.0%.
Conclusions:The posterior cornea significantly influences corneal astigmatism. Its power value and orientation are much more stable with advancing age compared with anterior surface values.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.