In this paper, we studied additional filter and Ion chamber combinations to reduce patient dose without decreasing image quality in digital chest x-ray projection. The experiment set 125 kVp, 320 mA, AEC mode. Ion chamber sensors was divided by 4 cases of combinations, then, we measured patient dose and calculated organ dose using PCXMC. Also, physical image assessment using MTF was performed. As a results, The surface entrance dose and organ dose were the lowest when selecting both left and right Ion chamber sensors under the same conditions of additional filter. In image quality assessment, The spatial frequency scored 2.494 lp/mm which was highest when selecting both right and left Ion-chambers and 0.1 mmCu filter. And to conclude, to select both right and left Ion chamber sensors and 0.1 mmCu filter will help for acquiring good quality image as well as reducing patient dose.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.