A Comparison of Absorbed Doses to Water in Photon Beams in Japan Society of Medical Physics 01 and 12

Abstract: A comparison of absorbed doses to water at a calibration depth determined by Japan Society of Medical Physics (JSMP) 12 and 01 was conducted, using a farmer type ionization chamber. The absorbed dose to water calibration factor (ND,W,Q0) and beam quality conversion factor (kQ,Q0) for JSMP 12 were smaller than the absorbed dose to water calibration factor and beam quality conversion factor for JSMP 01. Differences in absorbed doses at a calibration depth were -0.78% for 6 MV photon beam and -0.94% for 10 MV pho… Show more