A comparison of the efficacy of multiple ultraviolet light room decontamination devices in a radiology procedure room

Abstract: ObjectiveTo evaluate the efficacy of multiple ultraviolet (UV) light decontamination devices in a radiology procedure room.DesignLaboratory evaluation.MethodsWe compared the efficacy of 8 UV decontamination devices with a 4-minute UV exposure time in reducing recovery of methicillin-resistant Staphylococcus aureus (MRSA), vancomycin-resistant Enterococcus (VRE), and Clostridium difficile spores on steel disk carriers placed at 5 sites on a computed tomography patient table. Analysis of variance was used to com… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

1
27
0

Year Published

2019
2019
2021
2021

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 25 publications
(28 citation statements)
references
References 22 publications
1
27
0
Order By: Relevance
“…Compared with the vertical tower low-pressure mercury devices, equal or greater reductions of the pathogens were achieved by 3 nonstandard low-pressure mercury devices that included either adjustable bulbs that could be oriented directly over the exam table or 3 vertical towers operated simultaneously. 4 Our results in achievable reduction factors in a test environment simulating real-life conditions of manual application are comparable to those of Cadnum et al However, 60 seconds were needed with our surfaces compared to the fast reduction achieved on agar plates by Petersson et al 3 and in our pretest. The large variability of log 10 unit reduction between tests and devices might be due to unavoidable differences in angulation of the device toward the surface, changing reflection artefacts, and/or the different surfaces, which are to be expected in a reallife environment.…”
Section: Discussionsupporting
confidence: 85%
See 2 more Smart Citations
“…Compared with the vertical tower low-pressure mercury devices, equal or greater reductions of the pathogens were achieved by 3 nonstandard low-pressure mercury devices that included either adjustable bulbs that could be oriented directly over the exam table or 3 vertical towers operated simultaneously. 4 Our results in achievable reduction factors in a test environment simulating real-life conditions of manual application are comparable to those of Cadnum et al However, 60 seconds were needed with our surfaces compared to the fast reduction achieved on agar plates by Petersson et al 3 and in our pretest. The large variability of log 10 unit reduction between tests and devices might be due to unavoidable differences in angulation of the device toward the surface, changing reflection artefacts, and/or the different surfaces, which are to be expected in a reallife environment.…”
Section: Discussionsupporting
confidence: 85%
“…A mobile, nonchemical device would therefore be a valuable alternative for a targeted surface disinfection. Cadnum et al 4 compared multiple UV decontamination devices in a radiology suite. Moreover, 4 standard, vertical-tower, low-pressure, mercury devices achieved reductions of VRE or MRSA ≥2 log 10 units and of C. difficle at~1 log 10 unit, whereas a pulsed-xenon device resulted in less reduction of the pathogens.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…UV-C light reduces contamination of high-touch clinical surfaces, yet more studies are needed to test the comparative efficacy of UV-C devices in real-world clinical environments. [5][6][7] We tested the efficacy of 2 UV-C devices in clinical hyperbaric chambers. The use of the Tru-D vegetative cycle and the 5-minute Moonbeam cycle resulted in similar reductions in MRSA; both resulted in significantly larger reductions than the manufacturer's recommended 3-minute Moonbeam-3 cycle.…”
Section: Discussionmentioning
confidence: 99%
“…Though utilization of UV light as a terminal cleaning procedure involving rotor wing aircraft is uncommon, there is promising literature involving UV light utilization in ambulances, radiology suites, and operating rooms 22,23 . This step of the decontamination process was necessary to provide additional protection for our flight crews and for subsequent patients.…”
Section: Development Of Rotor Wing Clinical and Operation Guidelines mentioning
confidence: 99%