Abstract:Robots can protect healthcare workers from being infected by the COVID-19 and play a role in throat swab sampling operation. A critical requirement in this process is to maintain a constant force on the tissue for ensuring a safe and good sampling. In this study, we present the design of a disposable mechanism with two non-linear springs to achieve a 0.6 N constant force within a 20 mm displacement. The nonlinear spring is designed through optimization based on Finite Element Simulation and Genetic Algorithm. … Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.