Abstract:2014 Un thermomètre à poudre de CMN dilué (Ce0,03La0,97) et un thermomètre à poudre de platine destinés à mesurer la chaleur spécifique de l'hélium-3 superfluide ont été réalisés. Leur comparaison suggère que la température magnétique T* du CMN dilué est reliée à la température absolue T par la relation T = T* + 0394 où 0394 =-0,16 ± 0,01 mK. Abstract. 2014 A powdered Ce0.03La0.97 magnesium nitrate thermometer has been developed to facilitate specific heat measurements in liquid 3He. Comparison with a platinum… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.