Background and study aims
Water drop adhesions (WDA) impair endoscopic view during gastrointestinal endoscopy. We developed a novel lens cleaner designed using two types of harmLess surfactants and it is reported to be useful for preventing lens cloudiness during colorectal ESD. In the current study, we examined the ability of it for preventing and removing WDA.
Patients and methods
During laboratory experiments, the cleaner (Cleash; Fujifilm Co., Tokyo, Japan and Nagase Medicals Co., Hyogo, Japan) was applied to the endoscopic lens and an air/water device (AWD) (water 200 mL, dimethicone 1 mL, Cleash 1 mL). The endoscope was submerged in water 100 times for 5 cycles. Rates of WDA were calculated for various groups (lens and AWD with or without Cleash) and compared to a normal cleaner (SL cleaner). During clinical research, 30 colonoscopies and 30 esophagogastroduodenoscopies were analyzed. For the Cleash group, the cleaner was applied to both lens and AWD. The numbers of WDA and WDA with non-rapid removal were calculated, compared to those of the SL cleaner group.
Results
The mean WDA rate for the Cleash setting (lens: Cleash; AWD: Cleash) was 11.0 %, which was significantly lower than other settings (lens: SL cleaner; AWD: water, 31.0 %;
P
< 0.001) (lens: Cleash; AWD: water, 19.0 %;
P
< 0.001). Clinical research of colonoscopies indicated that the numbers of WDA (number/15 sec) and WDA with non-rapid removal were 0.38 and 0.17 for the Cleash group and 0.91 and 0.46 for the SL cleaner groups (
P
< 0.001,
P
< 0.001). For esophagogastroduodenoscopies, the results were 0.47 and 0.24 for the Cleash group and 0.54 and 0.42 for the SL cleaner group (
P
= 0.72,
P
= 0.018).
Conclusion
A clear and beautiful image without WDA is useful not only for routine endoscopy but also, more importantly, for magnifying endoscopy and other endoscopic treatments. The use of Cleash to lens and AWD showed positive results for preventing and removing WDA during laboratory experiments and clinical research involving CS. Additionally, it also showed positive results for the removal of WDA during EGD.