A precise and accurate analytical method for determination of osmium isotope ratios at the 1–15 pg level by using a MC-ICP-MS equipped with sparging introduction and high-sensitivity discrete dynode-type ion-counting detectors

Abstract: Properly calibrated discrete dynode SEM ion-counting detectors and minimization of blank Os by an improved vial-cleaning method enables Os isotope analysis at the 1–15 pg level with high accuracy and precision.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

0
4
0

Year Published

2022
2022
2025
2025

Publication Types

Select...
5
1

Relationship

1
5

Authors

Journals

citations
Cited by 9 publications
(4 citation statements)
references
References 48 publications
0
4
0
Order By: Relevance
“…Thus, the memory effect due to the previous sample was quickly removed by the Ar gas flushing with less than 30 s. In this case, the memory effect due to analysis of a previous sample was quickly removed by Ar gas flushing for < 30 s. After the Os isotope measurement, the sample solutions were evaporated on a hotplate at 140 °C to remove the remaining Os. Rhenium was then purified using a two-step column separation procedure with anion exchange resin (Muromac AG 1-X8) following the methods of Nozaki et al 65 , Ohta et al 68 , and Morgan et al 70 . Rhenium isotope ratios were measured by MC–ICP–MS with solution introduction into the glass torch, using the Ir standard addition method for external mass bias correction by using 193 Ir/ 191 Ir = 1.68097 (Refs.…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations
“…Thus, the memory effect due to the previous sample was quickly removed by the Ar gas flushing with less than 30 s. In this case, the memory effect due to analysis of a previous sample was quickly removed by Ar gas flushing for < 30 s. After the Os isotope measurement, the sample solutions were evaporated on a hotplate at 140 °C to remove the remaining Os. Rhenium was then purified using a two-step column separation procedure with anion exchange resin (Muromac AG 1-X8) following the methods of Nozaki et al 65 , Ohta et al 68 , and Morgan et al 70 . Rhenium isotope ratios were measured by MC–ICP–MS with solution introduction into the glass torch, using the Ir standard addition method for external mass bias correction by using 193 Ir/ 191 Ir = 1.68097 (Refs.…”
Section: Methodsmentioning
confidence: 99%
“…Rhenium isotope ratios were measured by MC–ICP–MS with solution introduction into the glass torch, using the Ir standard addition method for external mass bias correction by using 193 Ir/ 191 Ir = 1.68097 (Refs. 68 , 71 ). The Re blank was 0.44 pg, and the Os blanks were 0.12 and 0.09 pg, with 187 Os/ 188 Os ratios of 0.157 and 0.221, respectively.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations