A thermoelectrically stabilized aluminium acoustic trap combined with attenuated total reflection infrared spectroscopy for detection of Escherichia coli in water

Abstract: Acoustic trapping is a non-contact particle manipulation method that holds great potential for performing automated assays. We demonstrate an aluminium acoustic trap in combination with attenuated total reflection Fourier-transform infrared...

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
3
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
2
1

Relationship

1
2

Authors

Journals

citations
Cited by 3 publications
(3 citation statements)
references
References 49 publications
0
3
0
Order By: Relevance
“…This geometry resulted in 14 active bounces (N) at a total effective pathlength (d e,tot = d e N) of 18.1 µm at 1100 cm -1 (n dia = 2.4, n sample = n water = 1.33) and a depth of penetration (d p ) of 1 µm [47]. The measurement setup comprised a custom built ATR mount and an aluminium flow cell previously reported by Freitag et al [48] and Baumgartner et al [23], which allowed the integration into the sample compartment of a Vertex 70 v FTIR spectrometer (Bruker, Germany). Spectra were collected with OPUS 8.1 software using a spectral resolution of 4 cm -1 , 32 scans per spectrum (double sided, forward-backward acquisition mode).…”
Section: Optical Setup For Ftir Spectroscopymentioning
confidence: 99%
See 1 more Smart Citation
“…This geometry resulted in 14 active bounces (N) at a total effective pathlength (d e,tot = d e N) of 18.1 µm at 1100 cm -1 (n dia = 2.4, n sample = n water = 1.33) and a depth of penetration (d p ) of 1 µm [47]. The measurement setup comprised a custom built ATR mount and an aluminium flow cell previously reported by Freitag et al [48] and Baumgartner et al [23], which allowed the integration into the sample compartment of a Vertex 70 v FTIR spectrometer (Bruker, Germany). Spectra were collected with OPUS 8.1 software using a spectral resolution of 4 cm -1 , 32 scans per spectrum (double sided, forward-backward acquisition mode).…”
Section: Optical Setup For Ftir Spectroscopymentioning
confidence: 99%
“…2. It is modified from the SIA system previously reported by Freitag et al [48], and consisted of a 10-port selection valve (VICI, Switzerland), as well as a Cavro XC syringe pump (Tecan, Switzerland) equipped with a 500 µL glass syringe (Tecan, Switzerland). The components were connected with PTFE tubing (VICI, Switzerland, O.D.…”
Section: Liquid Handling and Automationmentioning
confidence: 99%
“…The setup has been incorporated into a Fourier transform IR (FT-IR) spectrometer and was reported before. [45] Samples of the liquid headspace of the immobilized catalyst were taken and injecting it into the 20 μL flowcell utilizing a sequential injection analysis (SIA) system. For each measurement a total of 100 μL sample was injected.…”
Section: Introductionmentioning
confidence: 99%