1997
DOI: 10.1111/j.1365-2044.1997.184-az0321.x
|View full text |Cite
|
Sign up to set email alerts
|

Additional inspiratory resistance imposed by the laryngeal mask airway: in vitro versus in vivo comparison

Abstract: SummaryThe present study was designed: (a) to compare the additional inspiratory laryngeal mask airway (LMA) resistance measured in vitro during simulated ventilation and in vivo in five anaesthetised and mechanically ventilated patients; and (b) to evaluate the resistive pressure drop along the length of the LMA. After the differential pressure across the mask was measured, the pressure-flow relationship was characterised by Rohrer's equation and in vitro and in vivo resistance was calculated. Thereafter, the… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0
1

Year Published

1998
1998
2024
2024

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 12 publications
(3 citation statements)
references
References 23 publications
0
2
0
1
Order By: Relevance
“…The internal diameter of the syringe was 2 cm, which is approximately the size of an adult human trachea [ 9 , 10 ]. In the LMA group, a 35-mm- diameter plastic cylinder with a length of 120 mm was connected to the size 3 LMA to roughly simulate the insertion of LMA into the upper airway [ 11 ]. The balloon pressure was measured using VBM Balloon Pressure Gauge Handheld Pressure Pump Tracheal Intubation Pressure Gauge ( German, Hxh-11 Hxh Ce ).…”
Section: Methodsmentioning
confidence: 99%
“…The internal diameter of the syringe was 2 cm, which is approximately the size of an adult human trachea [ 9 , 10 ]. In the LMA group, a 35-mm- diameter plastic cylinder with a length of 120 mm was connected to the size 3 LMA to roughly simulate the insertion of LMA into the upper airway [ 11 ]. The balloon pressure was measured using VBM Balloon Pressure Gauge Handheld Pressure Pump Tracheal Intubation Pressure Gauge ( German, Hxh-11 Hxh Ce ).…”
Section: Methodsmentioning
confidence: 99%
“…Katz, et al (1985) first simulated spontaneous breathing with a mechanical device while measuring continuous positive airway pressure using a two-chambered lung model. Righini, et al (1997) used a one-lung model when measuring the additional inspiratory resistance imposed by the Laryngeal Mask Airway. Austin, et al (2000) used a two-chambered test lung to measure work of breathing characteristics of portable ventilators.…”
Section: Introductionmentioning
confidence: 99%
“…후두마스크는 기관 내튜브에 비하여 여러 가지 장점을 갖는다 13). 후두마스크는 장치 자체의 기도저항이 기관내튜브에 비해 작다 14,18). 낮은 기도저항은 기관내튜브에 비해 호흡일(work of breathing)을 감소시키고19) 동적 유순도(dynamic compliance)를 높인다.참 고 문 헌…”
unclassified