2019
DOI: 10.1213/xaa.0000000000001045
|View full text |Cite
|
Sign up to set email alerts
|

Adjustable Pressure Limit Valve Failure During Intraoperative Care: A Case Report

Abstract: One of the many safety features of modern day anesthesia machines is the adjustable pressure limiting (APL) valve. This device regulates pressure within the anesthesia circuit during manual ventilation with the anesthesia bag. We report an unusual case where a crack in the APL valve allowed release of pressure from within the circuit resulting in ineffective bag-valve-mask ventilation of an infant. The appropriate steps to prevent such issues are reviewed, and an algorithm to quickly identify such intraoperati… Show more