Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects

Abstract: Japan Clinical Trials Register; number: UMIN000011544.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

6
38
2
1

Year Published

2017
2017
2024
2024

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 56 publications
(47 citation statements)
references
References 27 publications
6
38
2
1
Order By: Relevance
“…In our prospective study of healthy older subjects with a mean age of 72 years, Icare TA01i gave lower mean IOP compared to GAT (−2.46±2.10 mmHg; 95% LOA was between −8.25 and 1.66 mmHg). 14 These age-related IOP changes noted using the rebound tonometer were first reported by González-Méijome et al 20 The mean patient age was 45 years (range 18–85 years). They found a negative correlation between age and IOP in the nasal and temporal cornea, but not in the corneal center.…”
Section: The Characteristics Of Each Device and Iop Valuesmentioning
confidence: 60%
See 2 more Smart Citations
“…In our prospective study of healthy older subjects with a mean age of 72 years, Icare TA01i gave lower mean IOP compared to GAT (−2.46±2.10 mmHg; 95% LOA was between −8.25 and 1.66 mmHg). 14 These age-related IOP changes noted using the rebound tonometer were first reported by González-Méijome et al 20 The mean patient age was 45 years (range 18–85 years). They found a negative correlation between age and IOP in the nasal and temporal cornea, but not in the corneal center.…”
Section: The Characteristics Of Each Device and Iop Valuesmentioning
confidence: 60%
“…The repeatability of Icare PRO, recorded as ICC (three sets of IOP readings), was 0.881 14 and 0.863 4 in the sitting position, and 0.656 4 in the supine position.…”
Section: The Characteristics Of Each Device and Iop Valuesmentioning
confidence: 97%
See 1 more Smart Citation
“…Second, IOP was measured by noncontact tonometry. Although a good inter-device agreement between Goldmann tonometry and noncontact tonometry has been reported, [27,28] Goldmann tonometry is still the gold standard in IOP assessment. However, Goldmann readings also have their limitations with cloudy, edematous corneas, and Fig.…”
Section: Discussionmentioning
confidence: 99%
“…Intraocular pressure was measured by corneal rebound tonometry directly before every SD-OCT examination was initiated (Icare tonometer TA01i; Icare Finland Oy, Vantaa, Finland). [25][26][27]…”
Section: Methodsmentioning
confidence: 99%