2020
DOI: 10.1111/aos.14701
|View full text |Cite
|
Sign up to set email alerts
|

Agreement between rebound (Icare ic200) and applanation tonometry (Perkins) in patients with primary congenital glaucoma

Abstract: Purpose To examine agreement between intraocular pressure (IOP) measurements made using the rebound tonometer Icare ic200 (RT200) and the Perkins handheld applanation tonometer (PAT) in patients with primary congenital glaucoma (PCG). The impacts of several covariables on measurements using the two devices were also assessed. Materials and Methods Intraocular pressure measurements were made in a single session in 86 eyes of 86 patients with PCG (46 under anaesthesia, 40 in the office). The order was RT200 then… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

1
2
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(3 citation statements)
references
References 17 publications
1
2
0
Order By: Relevance
“…This is consistent with results obtained in our preliminary study using IC200 6. Mean difference between IOP Ic200 and Perkins IOP readings was 1.26 mm Hg, an upper limit of agreement of −8.06 mm Hg and a lower limit of agreement of 10.59 mm Hg.…”
Section: Discussionsupporting
confidence: 92%
See 2 more Smart Citations
“…This is consistent with results obtained in our preliminary study using IC200 6. Mean difference between IOP Ic200 and Perkins IOP readings was 1.26 mm Hg, an upper limit of agreement of −8.06 mm Hg and a lower limit of agreement of 10.59 mm Hg.…”
Section: Discussionsupporting
confidence: 92%
“…In our preliminary study, 6 systematic and proportional differences between IC200 and Perkins IOP readings were found. These differences were higher in high values of IOP, especially in PCG patients.…”
mentioning