2023
DOI: 10.1093/rpd/ncad127
|View full text |Cite
|
Sign up to set email alerts
|

Alanine/EPR dosimetry for mailed intercomparison at ocular proton therapy facilities—preliminary results for three centres for irradaitions at CCB IFJ PAN eyeline

Barbara Michalec,
Cinzia De Angelis,
Gabriela Foltyńska
et al.

Abstract: Quality control of therapeutic photon beams in the form of postal dose audits based on passive dosemeters is widely used in photon radiotherapy. On the other hand, no standardised dosimetry audit programme for proton centres has been established in Europe so far. We evaluated alanine/EPR dosimetry systems developed at the Istituto Superiore di Sanità (Italy), the Hasselt Universiteit (Belgium) and the Henryk Niewodniczanski Institute of Nuclear Physics Polish Academy of Sciences (Poland) for their applicabilit… Show more