Abstract:We present a 62-year-old patient with COVID-19 pneumonia on Veno-venous (VV) Extracorporeal Membrane Oxygenation (ECMO) with unique perturbations to pre and post oxygenator pressures due to fibrin deposition in despite being on a Heparin/Bivalirudin infusion and activated Partial Thromboplastin Time (aPTT) within therapeutic range of 60–80 seconds. On Day 8 of ECMO support, it was noticed that flows steadily decreased despite unchanged RPMs. Unlike typical blood flow to circuit pressure relationships, the circ… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.